000 01880nab a2200205 4500
005 20260520001830.0
008 260224s2010 xxu
245 0 0 _aOn the end-effect correction for Couette type oil-field direct-indicating viscometers for Newtonian and non-Newtonian fluids
260 _a
_b
_cmar. 2010
270 _a12/07/2010 ; 12/07/2010
300 _a10 p. ; 37-46
520 _aTranscripción del resumen del autor. Analysis and experimental results are presented which quantify the magnitude of end-effects for direct- indicating, narrow gap, rotating cylinder viscometers used extensively in oil- eld, for Newtonian and non- Newtonian uids. Such viscometers have already embedded at manufacturing stage an end-effect correction and give directly the shear stress which does not allow end-effect quanti cation in a straight forward manner. The presented analysis resolves this problem by estimating the embedded correction, which has been determined as 6.69% of the torque developed along the cylindrical part of the stationary bob of this viscometer. Experimental results with Newtonian and non-Newtonian uids show that there is no additional end-effect from the bottom of the stationary bob. However, there is additional end-effect contribution from the top section of the bob, which has not been taken into account in the design and manufacturing of the instrument. Its magnitude ranges from 5 to 6% for the Newtonian uid and the high shear rate range for the non- Newtonian uid, to 12% for the low shear rate range for the non-Newtonian uid tested, while the impact on drilling situations is also discussed.
581 _a1-2
773 0 _tJournal of Petroleum Science & Engineering
_g71
942 _cARTICULO
100 1 _aKelessidis, V.C.
_938573
100 1 _aMaglione, R.
_945600
100 1 _aBandelis, G.
_945601
999 _c174936